Specifications:
Test Frequencies: 500, 1000, 2000, 3000, 4000, and 6000 Hz. 8000 Hz
is optional.
Frequency Accuracy: Crystal controlled, less than 1% error at all
frequencies.
Frequency Sequence: Start either ear, 1000, 500, 1000 retest, 2000,
3000, 4000, 6000, (optional 8000). Second ear, 500, 1000, 2000, 3000,
4000, 6000 (and 8000 optional).
Intensity Range: From 0 dB to 90 dB Hearing Level (HL) in 5 dB steps.
Attenuator Linearity: Less than 0.75 dB error for any 5 dB step, less
than 1 dB error for any 10 dB step, less than 2 dB accumulated error
relative to the calibration level.
Tone Rise/Fall Times: 44/32 ms typical.
Test Paradigm: Modified Hugheson Westlake in the automatic mode with
fully manual override capability.
Testing Time: On a cooperative individual or a biological simulator,
approximately 5 ½ minutes (testing 8 kHz.).
Stimulus Characteristics
A - Pulse train is 1.2 seconds with 50% duty cycle (200 ms on and 200 ms
off) with three pulsed tone presentations.
B - Time between tone presentations is varied randomly between 1 and 2
seconds.
C - Patient response window is 1.8 seconds from the beginning of the
pulse train.
D - Pulse train terminates when response switch is depressed.
Error Warnings: Audible alert (when active) with visual display of
specific error encountered. The following error conditions are detected
and signaled.
A - Threshold not established at 1000 Hz, first ear.
B - Retest failure at 1000 Hz.
C - Patient responding when no tone is being presented.
D - Switch not released.
E - Failure to establish threshold within allotted time (pause and alert
optional at this error).
Audiometer Calibration: All audiometer calibration parameters will
meet ANSI S3.6 1996 Standard for Audiometers and OSHA 29 CFR 1910.95.
Output levels are calibrated through secured keyboard entry.
Safety: Listed to UL544 Standard for Medical and Dental Equipment and
CSA C22.2 No. 125-M1984 Environment Products/Health Care Technology.
Earphone: Telephonics Corporation TDH -49 earphones with MX-41/AR
cushions.
Power Requirements: 120 VAC, 60 Hz, 32 VA
Physical Dimensions: 2.75" high, 7.5" wide, 9" deep.
Net Weight: Unit, 2.6 lbs; Earphone, 1 lb; Transformer, 1.2 lbs.
Standard Equipment:
MI-5000B Audiometer with RS-232 output and internal talkover
Earphone Assembly
Patient Response Switch
Power Supply
Patch Cords (2)
Operator Manual
|